• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, CHINESE, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, CHINESE, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-45
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/04/2022
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) units iab loop is leaking.There was no patient harm reported.
 
Event Description
N/a.
 
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to investigate the issue.In order to fix the issue, the fse replaced the safety disk (0997-00-0985-15).The fse stated that the safety disk automatically tests the positive pressure leakage of 4mmhg, and the manual test of the x1 sensor and can obtain a positive pressure leakage of more than 10mmhg, where the maximum is 42mmhg.Also, the fse stated that the equipment alarm log cannot display the alarm.The fse performed all functional and safety checks to factory specifications.The unit was returned to the customer and cleared for clinical use.
 
Event Description
N/a.
 
Manufacturer Narrative
The following investigation was performed by a technician of the maquet national repair center (nrc) wayne, nj: the nrc received safety disk with the reported complaint of the safety disk failing the pressure test.The national repair center installed the safety disk into the cs 300 test fixture and tested the safety disk to factory specifications per the cs300 service manual.The nrc performed the safety disk leak test with passing results per factory specifications.K5 vacuum test 0 mmhg; k3 vacuum test 0 mmhg; pressure test -1 mmhg.Retaining the safety disk in the nrc per procedure number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, CHINESE, 220V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key15236770
MDR Text Key304922175
Report Number2249723-2022-02093
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108575
UDI-Public10607567108575
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 03/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-45
Device Catalogue Number0998-00-3023-45
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/04/2022
Initial Date FDA Received08/16/2022
Supplement Dates Manufacturer Received04/15/2023
03/05/2024
Supplement Dates FDA Received04/24/2023
03/05/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/08/2019
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-