DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, CHINESE, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-45 |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/04/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) units iab loop is leaking.There was no patient harm reported.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to investigate the issue.In order to fix the issue, the fse replaced the safety disk (0997-00-0985-15).The fse stated that the safety disk automatically tests the positive pressure leakage of 4mmhg, and the manual test of the x1 sensor and can obtain a positive pressure leakage of more than 10mmhg, where the maximum is 42mmhg.Also, the fse stated that the equipment alarm log cannot display the alarm.The fse performed all functional and safety checks to factory specifications.The unit was returned to the customer and cleared for clinical use.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
The following investigation was performed by a technician of the maquet national repair center (nrc) wayne, nj: the nrc received safety disk with the reported complaint of the safety disk failing the pressure test.The national repair center installed the safety disk into the cs 300 test fixture and tested the safety disk to factory specifications per the cs300 service manual.The nrc performed the safety disk leak test with passing results per factory specifications.K5 vacuum test 0 mmhg; k3 vacuum test 0 mmhg; pressure test -1 mmhg.Retaining the safety disk in the nrc per procedure number.
|
|