• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL (SINGAPORE) BD HYPOINT¿ HYPODERMIC NEEDLE; HYPODERMIC SINGLE LUMEN NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL (SINGAPORE) BD HYPOINT¿ HYPODERMIC NEEDLE; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Model Number 300253
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.Medical device expiration date: unknown.Device manufacture date: unknown.Investigation summary: the customer issued a complaint for broken hypoint needle during use by end user in the market.No sample was provided to bd medical ¿ pharmaceutical system (bdm-ps) for analysis.Bdm-ps performed a batch history record¿s review (bhr) including a review of all data collected during in process and quality inspections.The batch involved in this complaint meets all acceptable quality levels (aql¿s), was manufactured, and released according to applicable procedures and specifications.As no sample was provided to bdm-ps for analysis and no additional information is available, the 8d team leader is not able to delve deeper into the investigation and correlate this symptom with a potential cause linked to bdm-ps hypoint manufacturing process.Based on the description bd were not able to assign a quality criteria that is related to the issue.However even taking the most severe quality limit defined in the agreed specification the occurrence is within the specification and no further action will be assigned and report will not be updated on receipt of additional sample information.
 
Event or Problem Description
It was reported that the bd hypoint¿ hypodermic needle broke off while attempted to inject firazyr 30mg during the infusion.The following information was provided by the initial reporter: "patient informed that they went to infuse there firazyr 30mg pfs and the needle broke as they went to inject.No injury to patient.They then put this into the sharps bin and used their spare firazyr 30mg pfs.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD HYPOINT¿ HYPODERMIC NEEDLE
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer (Section G)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15236969
Report Number8041187-2022-00470
Device Sequence Number20006863
Product Code FMI
UDI-Device Identifier00382903002535
UDI-Public00382903002535
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K070440
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number300253
Device Catalogue Number300253
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/19/2022
Initial Report FDA Received Date08/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-