Issue description on (b)(6)2022, a customer from belgium notified biomérieux of legionella contamination of nucl.Magn.Silica 384t - 280133, lot number z013ad1ms, expiration date: 28 jan 2023) leading to a delayed result when testing patient sample.The result were not given to the physician that leading to a delay in rendering results.However, the length of the delay is not known yet.There is no indication or report from the laboratory that the issue led to any adverse event related to patient's state of health.A biomérieux internal investigation has been initiated.
|
An internal investigation was performed following notification from a customer in belgium of bacterial contamination when using nuclisens® magnetic silica (ref.(b)(4), lot number z013ad1ms, expiration date: 28 jan 2023).The investigation confirmed that this lot was not contaminated.A contamination issue was confirmed with specific lots of nuclisens magnetic silica, for which fsca 5690 was issued.However, the silica lot involved in this instance, z013ad1ms, is not part of the listed impacted lots in fsca 5690.Testing was done on retained kits from the manufacturing site and also on raw material for batches currently released on the field at that time.The investigation demonstrated that the contamination comes from silica raw material from a supplier.The silica raw material produced internally are not impacted by the issue.List of incriminated lots: z012me1ms, z012mf1ms, z012mh1ms, z012mk1ms, z012ml1ms, z012mg1ms, z012ne1ms, z012nd1ms, z012nc1ms, z012nb1ms, z013af1ms, z013ag1ms, z013ah1ms, z013ak1ms, z013al1ms).All other batches have been tested by the investigation r&d team.These lots, including the lot involved in this complaint, were conforming and not concerned by the contamination issue.In addition, no more contamination has been observed using these other silica batches by some complaint customers on incriminated batches.Some customers may have observed contamination with these uncontaminated silica lots due to: a difference of sensitivity in the pcr test compared to tests done by investigation r&d team.A residual environmental contamination in customer lab.Or possibly the use of pooled batches within contaminated ones.(pooling is not part of good practices) for these customers, the recommendation is to discard all opened reagents, (ic, silica), to perform a decontamination of the lab and to run contamination/blank tests after.
|