Model Number M00568211
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/20/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that an endovive standard peg push method tube was used during a percutaneous endoscopic gastrostomy placement procedure on (b)(6) 2022.During the procedure, when physician tried to pull through the stomach, the peg tube broke in half.The procedure was completed with a new endovive standard peg push method.There were no reported patient complications as a result of this event.
 
Additional Manufacturer Narrative
(b)(4).The returned endovive standard peg tube was analyzed however the rigid tubing was not returned.No damage was observed with any returned part.Product analysis found the peg tube separated from the dilator at the barbed connector joint.The reported complaint was confirmed.Based on the condition of the returned device, engineers determined that the problem observed is likely that operational factors, such as user technique/handling, patient anatomy, and the size of the incision site that the tube was being pulled through, contributed to the reported complaint.Boston scientific has determined the most probable cause of this complaint is adverse event related to procedure.It is most likely that the adverse event occurred during the procedure and the device had no influence on event which led to the reported event.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOVIVE STANDARD PEG KIT
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15237415
Report Number3005099803-2022-04520
Device Sequence Number6293427
Product Code KNT
UDI-Device Identifier08714729285205
UDI-Public08714729285205
Combination Product (Y/N)Y
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K031538
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/22/2022
Device Model NumberM00568211
Device Catalogue Number6821
Device Lot Number0028843475
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/01/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/21/2022
Initial Report FDA Received Date08/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/10/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-