| Model Number |
S-60-120-120-P6 |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.It is possible that interactions with the anatomy and/or a coagulation of blood and contrast on the guide wire resulted in the reported difficult to remove; however, this cannot be confirmed.The investigation determined a conclusive cause for the reported difficulties cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.The command 18 guidewire referenced is filed under a separate medwatch report number.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a lesion located in the superficial femoral artery.A supera stent was successfully deployed, however, when performing a wire exchange with a command 18 guidewire, wire access was lost due to resistance with the guidewire and supera delivery system.The delivery system was removed and the vessel re-wired, without further issue or intervention.The stent was well placed and the lesion well treated.There was no reported adverse patient effect and no clinically significant delay in the procedure.A command 014 wire was successfully used to continue the procedure.No additional information was provided.
|
| |
|