MENTOR TEXAS MENTOR SILTEX ROUND MODERATE PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
|
Back to Search Results |
|
| Model Number |
3542615 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
06/22/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a female patient who underwent breast augmentation with a mentor siltex round moderate profile 150cc saline prosthesis experienced left sided deflation post procedure.As a result, explant was performed on an unknown date post procedure.No additional event details are available at this time.If additional event details become available in the future, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: upon visual evaluation of the image provided in the complaint, the implant was observed deflated.As the product involved in this complaint was not received, the device couldn´t be analyzed according to our procedures.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.Reason for device explant and/or reoperation: deflation.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was received on (b)(6) 2022.The implant date and explant date were obtained.The device was implanted on (b)(6) 2004.The device was explanted on (b)(6) 2022.The mentor failure analysis lab received the device for evaluation (b)(6) 2022.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation of the returned device was completed by the failure analysis lab on september 30, 2022.Device evaluation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection and leak testing of the returned device.Visual analysis of the returned sample revealed that no damage or anomalies were observed on the breast implant.Leak testing was performed according to the mentor procedures, and it revealed a tear within the siltex cracking posterior view measuring approximately 0.3 cm.The evaluation determined that the cause of the deflation is consistent with normal wear.Siltex cracking is defined as a material separation occurring in the siltex layer and can eventually progress through the shell of siltex devices.Siltex cracking is ultimately a result of high stresses.It should be noted that as part of the mentor¿s quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|