Model Number TV-AB3480-E
Medical Device Problem Codes Appropriate Device Problem Term/Code Not Available (3191); Migration (4003)
Health Effect - Clinical Codes Cardiac Arrest (1762); Shock (2072); Ruptured Aneurysm (4436)
Date of Event 08/04/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.
 
Event or Problem Description
The patient was initially treated for an abdominal aortic aneurysm (aaa) with implant of the ovation prime abdominal aortic aneurysm stent system.Approximately seven and a half (7.5) years post initial procedure, the patient presented emergently with a ruptured aneurysm and in shock.The patient was successfully resuscitated, and a potential type i endoleak was detected.The physician elected to treat the patient by implanting an afx vela infrarenal proximal extension on (b)(6) 2022.One (1) day post secondary intervention, the patient experienced cardiac arrest, but was treated and successfully returned to stable condition.
 
Event or Problem Description
The patient was initially treated for an abdominal aortic aneurysm (aaa) with implant of the ovation prime abdominal aortic aneurysm stent system.Approximately seven and a half (7.5) years post initial procedure, the patient presented emergently with a ruptured aneurysm and in shock.The patient was successfully resuscitated, and a potential type i endoleak was detected.The physician elected to treat the patient by implanting an afx vela infrarenal proximal extension on (b)(6) 2022.One (1) day post secondary intervention, the patient experienced cardiac arrest, but was treated and successfully returned to stable condition.Additional information received per clinical assessment indicating that implant migration (12mm) of the main body also occurred at time of reported event.In addition, the reported type i endoleak was confirmed to be a type ia endoleak.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the reported type ia endoleak, ruptured abdominal aortic aneurysm and additional endovascular procedure are confirmed.The reported hemodynamic instability, cardiac arrest and patient resuscitation are unconfirmed.This is moderately consistent with the reported adverse event/incident.The clinical evaluation also shows reasonable evidence to suggest 12mm migration of the main body occurred that was not included in the event as reported.This finding was discovered during an examination of ct (computed tomography) scan dated (b)(6) 2022.Procedure-related harms, device, user, procedure, or anatomy relatedness of this event could not be determined with the medical records available for review.The final patient status was reported as in stable condition.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Corrections: b5 describe event or problem.G3 awareness date.H6 investigation finding codes; remove 3233.H6 investigation conclusion codes; remove 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OVATION PRIME
Common Device Name
MAIN BODY
Manufacturer (Section D)
ENDOLOGIX SANTA ROSA
3910 brickway blvd
santa rosa CA 95403
Manufacturer (Section G)
ENDOLOGIX SANTA ROSA
3910 brickway blvd
,
santa rosa CA 95403
Manufacturer Contact
gary kirchgater
3910 brickway blvd
,
santa rosa, CA 95403
8009832284
MDR Report Key15238119
Report Number3008011247-2022-00092
Device Sequence Number17368471
Product Code MIH
UDI-Device IdentifierM701TVAB3480E1
UDI-Public+M701TVAB3480E1/$$3171104FS081414371
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P120006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 08/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/04/2017
Device Model NumberTV-AB3480-E
Device Catalogue NumberTV-AB3480-E
Device Lot NumberFS081414-37
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/04/2022
Supplement Date Received by Manufacturer10/04/2022
Initial Report FDA Received Date08/16/2022
Supplement Report FDA Received Date10/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/01/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OVATION PRIME FILL POLYMER, LN FF101714-01.; OVATION PRIME ILIAC LIMB, SN (B)(6).; OVATION PRIME ILIAC LIMB, SN (B)(6).
Outcome Attributed to Adverse Event Required Intervention;
Patient Age81 YR