• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILLIAM COOK EUROPE GUNTHER TULIP NAVALIGN FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM COOK EUROPE GUNTHER TULIP NAVALIGN FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Model Number G52917
Medical Device Problem Code Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Serious Injury
Event or Problem Description
Description of event according to initial reporter: the white plastic hub that houses the safety and releases mechanism was not aligned properly.They stated it was a very small misalignment.They realigned it and continued with the procedure.When trying to deploy, the filter would not disengage from the release mechanism.They then tried to re-sheath it and couldn¿t cover the hook on the filter.On the way out the filter got snagged and released in the femoral vein.The filter was off of the device but the feet were not deployed.The filter then had to be snared from a jugular vein approach.Patient outcome: no harm came to the patient.
 
Additional Manufacturer Narrative
Manufacturer ref# (b)(4).Pma/510(k) k211874.Investigation still in progress.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Manufacturer ref# (b)(4).Summary of investigational findings: it was reported that the filter would not disengage when attempted to be deployed.The filter was attempted to be resheathed but the physician couldn´t cover the filter hook.When the physician attempted to remove the device, the filter got released in the femoral vein.The filter had to be removed from a jugular approach.No harm to the patient reported.It was assessed that because any discovered non-conformances were properly dispositioned before qc release, there is evidence that the dhr was fully executed.In addition, no other lot related complaints have been received from the field.It is therefore concluded that there is no evidence that nonconforming product exists in house or in field.According to the instruction for use potential adverse events that may occur include, but are not limited to, the following: filter malpositioning there are adequate controls in place to ensure the device was manufactured to specifications.No device or images were provided for the investigation and with the information provided an exact cause for this event cannot be established.However, a possible cause is that the manipulation with the femoral filter introducer and the attempt to resheath the device contributed to the event, but this is purely speculations.Cook medical will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUNTHER TULIP NAVALIGN FEMORAL VENA CAVA FILTER SET
Common Device Name
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
Manufacturer (Section G)
WILLIAM COOK EUROPE
sandet 6, dk-4632
bjaeverskov
Manufacturer Contact
lissi walmann
sandet 6, dk-4632
bjaeverskov 
MDR Report Key15238426
Report Number3002808486-2022-00922
Device Sequence Number20447244
Product Code DTK
UDI-Device Identifier10827002529172
UDI-Public(01)10827002529172(17)241125(10)E4191246
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/25/2024
Device Model NumberG52917
Device Catalogue NumberIGTCFS-65-1-FEM-TULIP
Device Lot NumberE4191246
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/09/2022
Supplement Date Received by Manufacturer10/28/2022
Initial Report FDA Received Date08/17/2022
Supplement Report FDA Received Date10/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-