| Model Number |
SD980.016 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Difficulty Chewing (4522)
|
| Date of Event |
11/19/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device met specifications.Patient's bone did not heal in a timely manner as it is still not healed after 8 months.Replacement surgery planned for (b)(6) 2022.X-ray images of the broken plate provide sufficient information.
|
| |
|
Event or Problem Description
|
|
Left bsso plate broke after approximately 8 months and will be replaced with new bsso plate.
|
| |
|
Additional Manufacturer Narrative
|
|
Both bsso plates met specifications.Patient's bone did not heal in a timely manner as it is still not healed after 8 - 10 months.H3: x-ray images and pictures of the broken plates provide sufficient information.
|
| |
|
Event or Problem Description
|
|
Left bsso plate broke after approximately 8 months and will be replaced with new bsso plate.Additionally, the right bsso plate broke after approximately 8 -10 months and will be replaced with new bsso plate.Revision surgery scheduled for (b)(6) 2022.
|
| |
|