MAUDE Adverse Event Report: MATERIALISE NV TRUMATCH; TRUMATCH MIDFACE/MANDIBLE - TITANIUM 3D PRINTED PLATE FOR MANDIBLE, MINI

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MATERIALISE NV TRUMATCH; TRUMATCH MIDFACE/MANDIBLE - TITANIUM 3D PRINTED PLATE FOR MANDIBLE, MINI Back to Search Results
Model Number SD980.016
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Difficulty Chewing (4522)
Date of Event 11/19/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device met specifications.Patient's bone did not heal in a timely manner as it is still not healed after 8 months.Replacement surgery planned for (b)(6) 2022.X-ray images of the broken plate provide sufficient information.
 
Event or Problem Description
Left bsso plate broke after approximately 8 months and will be replaced with new bsso plate.
 
Additional Manufacturer Narrative
Both bsso plates met specifications.Patient's bone did not heal in a timely manner as it is still not healed after 8 - 10 months.H3: x-ray images and pictures of the broken plates provide sufficient information.
 
Event or Problem Description
Left bsso plate broke after approximately 8 months and will be replaced with new bsso plate.Additionally, the right bsso plate broke after approximately 8 -10 months and will be replaced with new bsso plate.Revision surgery scheduled for (b)(6) 2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUMATCH
Common Device Name
TRUMATCH MIDFACE/MANDIBLE - TITANIUM 3D PRINTED PLATE FOR MANDIBLE, MINI
Manufacturer (Section D)
MATERIALISE NV
technologielaan 15
leuven, vlaams-brabant 3001
BE  3001
Manufacturer (Section G)
MATERIALISE NV
technologielaan 15
leuven, vlaams-brabant 3001
BE   3001
Manufacturer Contact
jenny jones
technologielaan 15
leuven, vlaams-brabant 3001
BE   3001
MDR Report Key15238847
Report Number3003998208-2022-00017
Device Sequence Number4899764
Product Code JEY
UDI-Device Identifier05420060380167
UDI-Public05420060380167
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K170272
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberSD980.016
Device Catalogue NumberSD980.016
Device Lot NumberME21IMUHOD
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/01/2022
Supplement Date Received by Manufacturer09/09/2022
Initial Report FDA Received Date08/17/2022
Supplement Report FDA Received Date10/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age48 YR
Patient SexMale
-
-