• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ATLAS; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ATLAS; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number AT75186
Medical Device Problem Codes Material Frayed (1262); Peeled/Delaminated (1454); Difficult to Remove (1528); Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/27/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the during an angioplasty procedure, the catheter allegedly got struck in the wire and was allegedly unable to be removed.There was no reported patient injury.
 
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer.The investigation of the reported event is currently underway.(expiry date: 09/2024).
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one atlas pta dilatation catheter along with an unknown guide-wire with multiple kinks was returned for evaluation.Peeled pebax, fiber disturbance, and frayed fibers were noted on the balloon.No other specific anomalies were noted during the visual evaluation.On the functional testing of the returned device , the guide wire lumen was able to flush and the returned guide wire was able to insert and remove through in and out of the catheter without any issue.All the anomalies noted on microscopic observation.However, the investigation was confirmed for the identified peeled pebax, fiber disturbance, and frayed fibers as these anomalies were able to be observed on the returned balloon during the visual evaluation and microscopic observation.A definitive root cause for the reported guide wire removal failure and identified peeled pebax, fiber disturbance, and frayed fibers could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 09/2024), g3.H11: h6(result, conclusion).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the during an angioplasty procedure, the catheter allegedly got struck in the wire and was allegedly unable to be removed.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAS
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury 12804
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15239531
Report Number2020394-2022-00668
Device Sequence Number4693136
Product Code DQY
UDI-Device Identifier00801741062506
UDI-Public(01)00801741062506
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K120971
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 08/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAT75186
Device Catalogue NumberAT75186
Device Lot NumberGFFW2700
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/11/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/19/2022
Supplement Date Received by Manufacturer08/17/2022
Initial Report FDA Received Date08/17/2022
Supplement Report FDA Received Date08/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-