Catalog Number MC1005T
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/29/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a warehouse found an roi-c cage that is fractured and a plate that has a bent arm during inspection after they were returned from a hospital.Follow-up with the distributor determined the anchors were too small for the patient's vertebral body.The cage and anchors were removed and replaced with another larger size.Surgery was delayed approximately 30 minutes.This is report one of two for this event.
 
Additional Manufacturer Narrative
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.Reference report 3004788213-2022-00062.
 
Additional Manufacturer Narrative
Device evaluation the devices were returned in a contaminated state so only a visual inspection was able to be performed.The arms of the plate appear to be bent inwards and the cage was cracked around the plate point of entry.Potential root cause a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to excess force applied during removal.Dhr review the dhrs were reviewed.There are no indications of manufacturing issues which would have contributed to this event and the devices were likely conforming when they left zimvie¿s control.Device usage this device is used for treatment a follow-up report will be submitted if new information is received that changes the information provided in this report.
 
Event or Problem Description
It was reported that a warehouse found an roi-c cage that is fractured and a plate that has a bent arm during inspection after they were returned from a hospital.Follow-up with the distributor determined the anchors were too small for the patient's vertebral body.The cage and anchors were removed and replaced with another larger size.Surgery was delayed approximately 30 minutes.This is report one of two for this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROI-C SHORT ANCHORING PLATE
Common Device Name
ROI-C TITANIUM-COATED IMPLANT SYSTEM
Manufacturer (Section D)
LDR MÉDICAL
quartier europe de l¿ouest
5, rue de berlin
sainte-savine, na 10300
FR  10300
Manufacturer (Section G)
LDR MÉDICAL
quartier europe de l¿ouest
5, rue de berlin
sainte-savine, na 10300
FR   10300
Manufacturer Contact
kim martinez
10225 westmoor dr.
na
westminster, CO 80021
3035144809
MDR Report Key15240069
Report Number3004788213-2022-00061
Device Sequence Number4693154
Product Code OVE
UDI-Device Identifier03662663016971
UDI-Public(01)03662663016971(17)270301(10)289294/21
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K151934
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMC1005T
Device Lot Number289294/21
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/25/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/25/2022
Supplement Date Received by Manufacturer01/17/2023
Initial Report FDA Received Date08/17/2022
Supplement Report FDA Received Date01/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/29/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-