EXACTECH, INC. LOGIC CR TIB INSERT SLOPE+, SZ 4, 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Model Number |
02-012-48-4011 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994); Joint effusion (4849)
|
| Date of Event |
07/29/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported, approximately 5 years post op the initial tka, this male patient showed up to clinic with pain and recurrent effusions.Surgeon noted moderate poly wear and oxidation upon examining the insert.He also stated there was no significant osteolysis on imaging.New 11crc insert was implanted with good rom and stability.Patient left in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
Pending evaluation.Concomitant medical products: 02-010-03-0340, 4948624, logic cr femoral cem, right, sz 4.02-012-45-4040, 4949766, lgc tibial fit tray cem sz 4f / 4t.200-02-38, 4802134, three peg patella 38mm.204-34-02, 4906993, fluted stem extension 25l x 14 mm.204-70-00, 4934190, tibial stem ext.Screw.
|
| |
|
Additional Manufacturer Narrative
|
|
Section h10: (h3) the revision reported was likely the result of mechanical overload on the posterior medial edge of the tibial insert.Contributions from the posterior slope, rotational alignment and/or potential ligamentous/soft tissue imbalances to the sequence of events cannot be determined from the reported clinical information and the device was not available for evaluation.Additionally, a contributing factor to the severity of the wear and delamination may have been result of being packaged in a non-conforming vacuum bag for more than five years.Section h11: the following sections have corrected information: (h6) component code: 734, bearings.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b1, b2, d8, g1, h6: clinical code, investigation findings, investigation conclusions.The revision reported was likely the result of mechanical overload on the posterior medial edge of the tibial insert.Contributions from the posterior slope, rotational alignment and/or potential ligamentous/soft tissue imbalances to the sequence of events cannot be determined from the reported clinical information and the device was not available for evaluation.Additionally, a contributing factor to the severity of the wear and delamination may have been result of being packaged in a non-conforming vacuum bag for more than five years.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|