Model Number 204-22-11
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Implant Pain (4561)
Date of Event 07/27/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Pending evaluation.
 
Event or Problem Description
As reported, approximately 5 years post op the initial tsa, this female patient was brought back for left knee pain.The poly insert was removed with significant wear.Both implant were found to be well fixed.A new exactech size 2 15mm insert was implanted.
 
Additional Manufacturer Narrative
(h3) the revision reported may have been due to third body wear, patient-related conditions, instability, and/or malalignment between the femoral and tibial components, which led to prosthesis wear and pain.Additionally, a contributing factor to the reported wear may have been the result of being packaged in a non-conforming vacuum bag for more than five years.However, this cannot be confirmed because the tibial insert was not returned for evaluation and images of the explanted device were unable to be obtained.The following sections have corrected information: (h6) component code: 734, bearings.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The revision reported may have been due to third body wear, patient-related conditions, instability, and/or malalignment between the femoral and tibial components, which led to prosthesis wear and pain.Additionally, a contributing factor to the reported wear may have been the result of being packaged in a non-conforming vacuum bag for more than five years.However, this cannot be confirmed because the tibial insert was not returned for evaluation and images of the explanted device were unable to be obtained.The product associated with the reported event is within the scope of recall z-0019-2022; however, there is insufficient information to evaluate whether the subject issue of the recall was a cause or contributor to the reported event.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.".
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.Corrected information: b1, b2, h6 clinical code.Updated information: d10, g contact name, facility name/address, g2, g3.D10: (b)(6) - 02-012-45-2020 - lgc tibial fit tray cem sz 2f / 2t.(b)(6) - 200-02-32 - three peg patella 32mm.(b)(6) - 201-78-82 - collar fixation pin 2pk 40mm, quick release (b)(6) - 201-78-88 - 4" drill bit, mod.Hex 2-pk.(b)(6) - 234-02-02 - optetrak asy, ps cemented femoral, sz 2.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PS TIBIAL INSERTS SZ 2, 11MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
3523771140
MDR Report Key15240701
Report Number1038671-2022-00923
Device Sequence Number6305199
Product Code JWH
UDI-Device Identifier10885862048233
UDI-Public10885862048233
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K933610
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 01/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/02/2018
Device Model Number204-22-11
Device Catalogue Number204-22-11
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/29/2022
Supplement Date Received by Manufacturer02/20/2023
03/08/2024
01/29/2025
Initial Report FDA Received Date08/17/2022
Supplement Report FDA Received Date03/03/2023
01/28/2025
01/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/05/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.; UNK.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
-
-