| Model Number |
72204370 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
07/29/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that, during use, when inserting the proximal portion of the pip implant, the implant sheared in half and broke off while part of it was still in the patient and it was unable to be removed from the bone.Surgery was resumed, after a non-significant delay, with a change on the surgical technique; toe pinning.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the shipping information was provided; however, the subject device was not made available to the designated complaint unit and thus a physical product evaluation could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the material specification found that a statement (c of c) or evidence of conformance to material and processing specification must be provided.A clinical review states that there were no clinical factors found which would have contributed to the event.If a portion of the implant remains retained in the patient, it is made of a biocomposite material which is intended for implantation to allow for bone ingrowth.If the implant was retained, it is unknown if the broken piece will migrate and there is potential risk for tissue inflammation and/or pain.An analysis of the customer provided images found the anchor had become fractured.Factors that could have contributed to the reported event include an impact inconsistent with normal use or excessive force.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|
| |
|
Search Alerts/Recalls
|
|
|