Model Number CTF33
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/09/2022
Type of Reportable Event Malfunction
Event or Problem Description
Procedure performed: robotic hiatal hernia event description: rep was not present for the case.After the trocar was placed in the abdomen and fully seated, a camera was placed down the trocar then they heard a pop and the trocar snapped in half.There was no patient injury.The case was completed with a new device.Product is available for return.Additional information received on 10aug2022 via phone call from user facility: incomplete, (7/8) break occurred in middle of cannula after placing instrument through cannula.The break did not fragment.The trocar was inserted normally with no difficulty during insertion.Type of intervention: the case was completed with a new applied trocar.Patient status: no patient injury.
 
Additional Manufacturer Narrative
The event device is anticipated to be returned to applied medical for evaluation.A follow-up report will be provided upon completion of the investigation.
 
Additional Manufacturer Narrative
The event unit was not returned to applied medical for evaluation.As the event unit was not returned, applied medical is unable to determine if the event unit exhibited any non-conformances that could have contributed to the reported event.In the absence of the event unit, it is difficult to determine if the reported event was caused by a manufacturing non-conformance or circumstantial factors at the time of use.Applied medical has reviewed the details surrounding the event and is unable to determine the exact root cause of the event.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.
 
Event or Problem Description
Procedure performed: robotic hiatal hernia.Event description: rep was not present for the case.After the trocar was placed in the abdomen and fully seated, a camera was placed down the trocar then they heard a pop and the trocar snapped in half.There was no patient injury.The case was completed with a new device.Product is available for return.Additional information received on 10aug2022 via phone call from user facility: incomplete (7/8) break occurred in middle of cannula after placing instrument through cannula.The break did not fragment.The trocar was inserted normally with no difficulty during insertion.Type of intervention: the case was completed with a new applied trocar.Patient status: no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CTF33, 11X100 KII FIOS ZTHR 6/BX
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
Manufacturer Contact
aaron fulcher
22872 avenida empresa
rancho santa margarita, CA 92688
9497135765
MDR Report Key15241394
Report Number2027111-2022-00704
Device Sequence Number14648893
Product Code GCJ
UDI-Device Identifier00607915124212
UDI-Public(01)00607915124212(17)250525(30)01(10)1449728
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
K041795
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/25/2025
Device Model NumberCTF33
Device Catalogue Number101219301
Device Lot Number1449728
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/09/2022
Supplement Date Received by Manufacturer08/09/2022
Initial Report FDA Received Date08/17/2022
Supplement Report FDA Received Date11/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-