| Model Number |
M635TU70040 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Thrombosis/Thrombus (4440)
|
| Date of Event |
08/09/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that thrombosis occurred.A left atrial appendage (laa) closure procedure was being performed.Prior to the procedure the activating clotting time (act) was 94 seconds.Heparin was administered just prior to trans-septal crossing and act was then noted to be 277 seconds.A watchman truseal access system (was) was inserted.As the was was crossing the septum, a thrombus was noted on the tip of the was.The clot was aspirated and the procedure continued.A watchman flx closure device & delivery system was inserted and deployed.The closure device was successfully implanted.
|
| |
|
Event or Problem Description
|
|
It was reported that thrombosis occurred.A left atrial appendage (laa) closure procedure was being performed.Prior to the procedure the activating clotting time (act) was 94.Heparin was administered just prior to trans-septal crossing and act was then noted to be 277.A watchman truseal access system (was) was inserted.As the was was crossing the septum, a thrombus was noted on the tip of the was.The clot was aspirated and the procedure continued.A watchman flx closure device & delivery system was inserted and deployed.The closure device was successfully implanted.It was further reported that the thrombus aspiration was performed via a non-boston scientific catheter and there was visual confirmation that the thrombus was successfully aspirated.Intracardiac echocardiography (ice) was the imaging modality utilized to identify the thrombus.
|
| |
|
Additional Manufacturer Narrative
|
|
B5: information added to event summary.
|
| |
|
Search Alerts/Recalls
|