• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC INSERTION HANDLE F/HINDFOOT ARTHRODESIS NAIL-EX; GUIDE, SURGICAL, INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC INSERTION HANDLE F/HINDFOOT ARTHRODESIS NAIL-EX; GUIDE, SURGICAL, INSTRUMENT Back to Search Results
Model Number 03.008.007
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
Device report from canada reports an event as follows: it was reported that on 19-july-2022, the surgeon tried to attach the insertion handle to a hindfoot arthrodesis nail using the connecting screw, but it was very difficult to align the insertion handle with the nail.The insertion handle was damaged, but the surgeon eventually got the nail and handle connected.Surgery was delayed for 3 minutes, there was no reported adverse patient consequence.No further information is available.This report is for an insertion handle f/hindfoot arthrodesis nail-ex.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: a manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Part number: 03.008.007 lot number: 72p5339 manufacturing site: haegendorf release to warehouse date: 28 october 2020 the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that there was no damage or defects with the insrt h&le f/hindfoot arthro nail-ex, p/n: 03.008.007.A dimensional inspection was unable to be performed due to the complex geometry of the device.A functional test was not conducted since the mating device was not returned for examination.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed as the nsrt h&le f/hindfoot arthro nail-ex, p/n: 03.008.007 was found to have no damage or defects.No definitive root cause could be determined.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSERTION HANDLE F/HINDFOOT ARTHRODESIS NAIL-EX
Common Device Name
GUIDE, SURGICAL, INSTRUMENT
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
WERK HÄGENDORF
im bifang 6
haegendorf 4614
SZ   4614
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key15242021
Report Number2939274-2022-03258
Device Sequence Number11954292
Product Code FZX
UDI-Device Identifier10886982066855
UDI-Public(01)10886982066855
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03.008.007
Device Catalogue Number03.008.007
Device Lot Number72P5339
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/20/2022
Supplement Date Received by Manufacturer08/17/2022
Initial Report FDA Received Date08/17/2022
Supplement Report FDA Received Date09/12/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/28/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-