WRIGHTS LANE SYNTHES USA PRODUCTS LLC INSERTION HANDLE F/HINDFOOT ARTHRODESIS NAIL-EX; GUIDE, SURGICAL, INSTRUMENT
|
Back to Search Results |
|
| Model Number |
03.008.007 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Device report from canada reports an event as follows: it was reported that on 19-july-2022, the surgeon tried to attach the insertion handle to a hindfoot arthrodesis nail using the connecting screw, but it was very difficult to align the insertion handle with the nail.The insertion handle was damaged, but the surgeon eventually got the nail and handle connected.Surgery was delayed for 3 minutes, there was no reported adverse patient consequence.No further information is available.This report is for an insertion handle f/hindfoot arthrodesis nail-ex.This is report 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: a manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Part number: 03.008.007 lot number: 72p5339 manufacturing site: haegendorf release to warehouse date: 28 october 2020 the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that there was no damage or defects with the insrt h&le f/hindfoot arthro nail-ex, p/n: 03.008.007.A dimensional inspection was unable to be performed due to the complex geometry of the device.A functional test was not conducted since the mating device was not returned for examination.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed as the nsrt h&le f/hindfoot arthro nail-ex, p/n: 03.008.007 was found to have no damage or defects.No definitive root cause could be determined.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|