Model Number 1804300-28
Medical Device Problem Code Difficult to Advance (2920)
Health Effect - Clinical Code Perforation of Vessels (2135)
Date of Event 07/27/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The reported patient effect of perforation is listed in the xience sierra everolimus eluting coronary stent systems instructions for use (ifu) as a known patient effect of coronary stenting procedures.The investigation determined that the reported difficulties and treatment appear to be related to circumstances of the procedure as it is likely the device interacted with the difficult anatomy resulting in the reported difficult to advance and subsequent perforation of the vessel.The perforation was successfully treated with another device.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported that the procedure was to treat a moderately calcified distal obtuse marginal artery.The 3.0x28mm xience skypoint was implanted; however, resistance was felt with the anatomy and a perforation was noted at the distal end of the stent implant.A 2.8x19mm graftmaster was advanced with no issue and implanted, successfully sealing the perforation.There was no adverse patient sequela and no clinically significant delay.No additional information was provide.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15242324
Report Number2024168-2022-08838
Device Sequence Number14677462
Product Code NIQ
UDI-Device Identifier08717648233265
UDI-Public08717648233265
Combination Product (Y/N)Y
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/13/2023
Device Model Number1804300-28
Device Catalogue Number1804300-28
Device Lot Number1120941
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/27/2022
Initial Report FDA Received Date08/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient SexFemale
Patient Weight85 KG
-
-