• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. SURGICEL FIBRILLAR 2INX4IN(5.1CMX10.2CM); AGENT, ABSORBABLE HEMOSTATIC, NON-COLLAGEN BASED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. SURGICEL FIBRILLAR 2INX4IN(5.1CMX10.2CM); AGENT, ABSORBABLE HEMOSTATIC, NON-COLLAGEN BASED Back to Search Results
Model Number 1962
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/25/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The following information was requested, but unavailable: - how was the product purchased? - is there an indication of how the product was distributed? - is there any indication of the source? - based on the packaging, is there any indication of which market the original genuine product may have come from? - was the device used on a patient? if so, was there any patient consequence? photo investigation summary: the packaging was observed to be according to the process specifications and the product was manufactured by ethicon, the diversion product is confirmed, since the ethicon product was not distributed by an authorized affiliate.Investigation summary: the product was returned for evaluation.In a further investigation of the samples received, it was observed that certain carton shading appeared to be missing as compared to approved carton artwork.The inside foil pouch appearance appears shinier than the approved material appearance.In addition, it was noted timestamp was identical for all packages of the same line.Based on the information currently available, the counterfeit was identified during the investigation of the samples received.Event related to mw # 2210968-2022-06652, mw # 2210968-2022-06656, mw # 2210968-2022-06688.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
A suspected product inquiry of (b)(6) science trade co.,ltd., during an ipo investigation.China supply chain checked and there was no record of the absorbable hemostat products.Additional information was requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURGICEL FIBRILLAR 2INX4IN(5.1CMX10.2CM)
Common Device Name
AGENT, ABSORBABLE HEMOSTATIC, NON-COLLAGEN BASED
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key15242821
Report Number2210968-2022-06686
Device Sequence Number20006902
Product Code LMG
UDI-Device Identifier10705031003125
UDI-Public10705031003125
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
N12159
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1962
Device Catalogue Number1962
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/06/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/08/2022
Initial Report FDA Received Date08/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-