Model Number MM-410320
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 01/01/2021
Type of Reportable Event Serious Injury
Event or Problem Description
Patient called to report device migration, exchange from textured to smooth breast implants due to the patient¿s concern with the product.Healthcare professional reported right side rupture.The device has been explanted.
 
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
 
Event or Problem Description
Patient called to report bilateral device migration, exchange from textured to smooth breast implants due to the patient¿s concern with the product.Healthcare professional reported right side rupture.Device has been explanted.
 
Additional Manufacturer Narrative
Device photo analysis: visual analysis of the photographs identified red particles material was found on the shell/gel, wear abrasion, fold creases and one extended opening.Device analysis performed through photographs, due to the impossibility to perform microscopic analysis it is not possible to determine the most likely failure mode.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key15242991
Report Number9617229-2022-13953
Device Sequence Number15025972
Product Code FTR
UDI-Device Identifier10888628003637
UDI-Public10888628003637
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P040046
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/17/2019
Device Model NumberMM-410320
Device Catalogue NumberMM-410320
Device Lot Number2685758
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/22/2022
Supplement Date Received by Manufacturer08/31/2022
Initial Report FDA Received Date08/17/2022
Supplement Report FDA Received Date11/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/17/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR
Patient SexFemale
Patient Weight53 KG
-
-