Model Number M005ATS20000
Medical Device Problem Codes Device Alarm System (1012); Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/22/2022
Type of Reportable Event Malfunction
Event or Problem Description
Note: this report pertains to one of three devices used during the same procedure.Refer to manufacturer report # 3005099803-2022-04602 and 3005099803-2022-04603 for the associated device information.It was reported to boston scientific corporation that an alair bt catheter was used for severe asthma during a bronchial thermoplasty procedure performed on (b)(6) 2022.During the procedure, two catheters and three grounding pads were used.An error on the controller showed a problem with the grounding pad and catheter.The physician was able to confirm that the grounding pads were placed correctly and was making contact with the airway wall.The physician replaced the grounding pad and catheter but the same problem occurred.They tested the controller with multiple demo catheters and received the same error.The procedure was cancelled due to this event.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
The complainant was unable to provide the upn and lot number of the suspect device.Therefore, the manufacture and expiration dates are unknown.(b)(4).The device has not been received for analysis.Upon receipt and completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Additional Manufacturer Narrative
Block d4, h4: the complainant was unable to provide the upn and lot number of the suspect device.Therefore, the manufacture and expiration dates are unknown.Block h6: impact code f1001 is being used to capture the reportable problem of aborted/cancelled procedure.Block h11: block b5 has been updated based on additional information received august 26, 2022.
 
Event or Problem Description
Note: this report pertains to one of three devices used during the same procedure.Refer to manufacturer report # 3005099803-2022-04602, 3005099803-2022-04603 and 3005099803-2022-04673 for the associated device information.It was reported to boston scientific corporation that an alair bt catheter was used for severe asthma during a bronchial thermoplasty procedure performed on (b)(6) 2022.During the procedure, two catheters and three grounding pads were used.An error on the controller showed a problem with the grounding pad and catheter.The physician was able to confirm that the grounding pads were placed correctly and was making contact with the airway wall.The physician replaced the grounding pad and catheter but the same problem occurred.They tested the controller with multiple demo catheters and received the same error.The procedure was cancelled due to this event.There were no patient complications reported as a result of this event.***additional information received on august 26, 2022*** it was reported that the two bt catheters and one bt controller were used in the right lower lobe of the lungs during the first bronchial thermoplasty procedure.From the way it was described, the catheter basket was illuminating orange on the controller and the grounding pad icon was able to connect just fine and was well after three missed activations.Then, the physician was able to complete the procedure on august 16, 2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALAIR BRONCHIAL THERMOPLASTY RF CONTROLLER
Common Device Name
BRONCHIAL THERMOPLASTY SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
150 baytech dr.
san jose CA 95134
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15243069
Report Number3005099803-2022-04673
Device Sequence Number4572339
Product Code OOY
UDI-Device Identifier08714729796695
UDI-Public08714729796695
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P080032
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM005ATS20000
Device Catalogue NumberATS 200
Device Lot Number051205-001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/26/2022
Supplement Date Received by Manufacturer08/26/2022
Initial Report FDA Received Date08/17/2022
Supplement Report FDA Received Date09/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/09/2011
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-