| Model Number |
866112 |
| Medical Device Problem Code |
No Audible Alarm (1019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that the central monitor speaker failed.The device was reported to be in use on a patient, but no adverse event to the patient or user was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A follow-up report will be submitted upon completion of the investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
The philips technical consultant (tc) confirmed the event.The tc noted that there was a cable that had become disconnected.The tc reconnected the cable.The tc had ordered a system, but did not need it, and it was returned unused.This is a reportable malfunction.The cause of the event was a cable that was disconnected.The issue was resolved by realigning the cable.The device remained on site.
|
| |
|
Event or Problem Description
|
|
The customer reported that the central monitor speaker failed.The device was reported to be in use on a patient, but no adverse event to the patient or user was reported.
|
| |
|