• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HORIZON TI MED 6/CART 180/BOX; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HORIZON TI MED 6/CART 180/BOX; CLIP, IMPLANTABLE Back to Search Results
Model Number IPN914925
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Codes Wound Dehiscence (1154); Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 08/03/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Incident occurred on (b)(6) 2022, involved an 80-year-old patient treated for bypass surgery coronary left internal mammary on posterolateral and right internal mammary on anterior interventricular under circulation extra bodily.During the intervention the clips self-removed from the internal breast on which they had been applied.Consequence: significant blood loss of 400 ml.Proximal tear bleeding of the right internal mammary artery required a vascular filling , norepinephrine and then continuous follow up.During postoperative follow up, we observed a postoperative bleeding that stopped after giving 2g of fibrinogen to the patient.Patient is fine, he was discharged on (b)(6) 2022 and left the hospital.
 
Additional Manufacturer Narrative
Qn#(b)(4).The device has not been returned for investigation.
 
Additional Manufacturer Narrative
Qn#(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.However, it is necessary to receive the physical sample to perform a proper investigation, determine root cause & implement corrective actions.Part number is not being manufactured currently, however, another part number from the same family was used for the "verification of failure mode reported in the current manufacturing process" and was conducted as follows: 59 samples were taken from the current production, the samples were functionally inspected, and during the test issue reported "ligation failure-clip not reopened" was not observed in the current manufacturing process.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Event or Problem Description
Incident occurred on (b)(6) 2022, involved an 80-year-old patient treated for bypass surgery coronary left internal mammary on posterolateral and right internal mammary on anterior interventricular under circulation extra bodily.During the intervention the clips self-removed from the internal breast on which they had been applied.Consequence: significant blood loss of 400 ml.Proximal tear bleeding of the right internal mammary artery required a vascular filling , norepinephrine and then continuous follow up.During postoperative follow up, we observed a postoperative bleeding that stopped after giving 2g of fibrinogen to the patient.Patient is fine, he was discharged on (b)(6) 2022 and left the hospital.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HORIZON TI MED 6/CART 180/BOX
Common Device Name
CLIP, IMPLANTABLE
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
jasmine brown
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key15243121
Report Number3003898360-2022-00414
Device Sequence Number11954168
Product Code FZP
UDI-Device Identifier24026704696363
UDI-Public24026704696363
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K132658
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN914925
Device Catalogue Number002200
Device Lot Number73B2200445
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/16/2022
Supplement Date Received by Manufacturer08/18/2022
Initial Report FDA Received Date08/17/2022
Supplement Report FDA Received Date09/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization; Life Threatening;
Patient Age80 YR
Patient SexMale
-
-