| Model Number |
G30496 |
| Medical Device Problem Code |
Unsealed Device Packaging (1444)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfg: device evaluation has begun; however, a conclusion is not yet available.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
As reported, prior to use during an unknown procedure, a roadrunner uniglide hydrophilic wire guide's package was found to be damaged.The device was not used, and another device of the same type was used to complete the procedure.There were no adverse effects to the patient.Upon return and initial evaluation of the device, part of the seal on the side of the package was open.
|
| |
|
Additional Manufacturer Narrative
|
|
During the investigation, physical examination of the returned device revealed that the package's outer pouch was ripped and the seal was pulled up, leading to delamination of the plastic packaging/seal.There is evidence that the package was sealed and the device was manufactured to specification.Per a search of risk documentation and previous complaint data, there is no evidence to suggest that this device failure would be likely to cause or contribute to a death or a serious injury if the failure were to recur.Furthermore, there is no evidence that the device caused or contributed to a serious injury, as no life-threatening or permanently impairing injury was alleged, nor was intervention taken as a result of the device failure which would be required to prevent permanent impairment or damage to the patient.As such, the event is not reportable under fda 21 cfr part 803 as it does not meet the criteria for a death, serious injury, or reportable product malfunction.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
|
| |
|