• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH GSTRO FEED TBE W/Y PRT 20FR; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH GSTRO FEED TBE W/Y PRT 20FR; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number 8884720205
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.Device evaluation comment - the customer stated that the device will not be returned for evaluation because it was contaminated.
 
Event or Problem Description
Customer reports: nursing called to report that while sleeping, the gastrostomy probe came out of the patient.It was introduced again by general surgery to change the gastrostomy probe.No.20fr and lidocaine in jelly are requested.The balloon was filled with 15cc of sterile water.Additional information was provided on 08/17/22 stating that the catheter was in place for 5 days.The balloon tore with no fragmented pieces.The patient did not receive food for one night until the patient underwent surgery to insert the new tube and remains hospitalized, since the patient was already hospitalized.The complete catheter was changed, not just the balloon, and has remained in the patient until he is discharged by phd and so far no failure has been reported by the patient.
 
Additional Manufacturer Narrative
The device history record (dhr) was reviewed, and no abnormal process conditions were present during the manufacturing of the product that could have led to the reported customer complaint.No photographs or product sample/s were provided for evaluation therefore the reported condition could not be confirmed.The root cause and corrective actions could not be identified without a sample evaluation.The current processes and controls were found to be followed correctly, including packaging and all inspections performed on the product.No abnormal conditions were found that could trigger the reported condition.The current process is running according to product specifications meeting quality acceptance criteria.We will continue monitoring the process for any adverse trends that require immediate attention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GSTRO FEED TBE W/Y PRT 20FR
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key15243632
Report Number9612030-2022-03358
Device Sequence Number17286132
Product Code KNT
UDI-Device Identifier10884521007291
UDI-Public10884521007291
Combination Product (Y/N)N
Initial Reporter CountryCO
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 09/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8884720205
Device Catalogue Number8884720205
Device Lot Number2132736264
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/09/2022
Supplement Date Received by Manufacturer08/09/2022
Initial Report FDA Received Date08/17/2022
Supplement Report FDA Received Date09/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
Patient Weight53 KG
-
-