Model Number 83779
Medical Device Problem Codes Difficult to Remove (1528); Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluated by manufacturer: the device was returned for analysis.Visual inspection was performed.The introducer sheath, the pusher and the main coil were returned.The introducer sheath was inspected, and no damages were found.Twist lock was opened.The main coil was kinked and stretched.The pusher wire was kinked, stretched, and the interlocking arm was detached.Microscope inspection was performed.The interlocking arm of the pusher wire was detached.
 
Event or Problem Description
Reportable based on device analysis completed on (b)(6) 2022.It was reported that the coil was stuck inside the catheter.A 4mm x 15cm interlock coil was selected for use.During the procedure, it was noted that the coil was stuck in the middle of the catheter.The coil was removed together with the catheter.The procedure was completed with another of the same device.No patient complications were reported.However, device analysis revealed that the interlocking arm of the pusher wire was detached.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLOCK
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15243901
Report Number2124215-2022-29955
Device Sequence Number4784999
Product Code KRD
UDI-Device Identifier08714729764960
UDI-Public08714729764960
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K132578
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/03/2025
Device Model Number83779
Device Catalogue Number83779
Device Lot Number0028802585
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/18/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/27/2022
Initial Report FDA Received Date08/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/03/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MICRO CATHETER - 0.025INCH MEDICOS HIRATA LANTERN
Patient SexUnknown
-
-