MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. OPT DILATING 12X100 STABILITY; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. OPT DILATING 12X100 STABILITY; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY Back to Search Results
Model Number 2D12LT
Medical Device Problem Codes Crack (1135); Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/25/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during to an unknown procedure, the trocar when used fall apart.The surgery was delayed by 15 minutes but was completed successfully.There were fragments outside of the patient but it did not require any intervention.There were no patient consequences.
 
Additional Manufacturer Narrative
(b)(4).Batch: # unk.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested and the following was obtained: what is meant by ¿when used they fell apart¿? it splits when the shirt is worn was the damaged identified during use or right out of the package? no.Were the devices re-sterilized or were they opened directly out of the sterile pouch from the manufacture? i believe re-sterilized, they do not give lot information.How was the surgery completed? yes.Were pieces of the trocar left within the patient¿s abdomen? none.If so, how were those fragments generated during the procedure? what type of intervention was performed to remove the fragments from the patient? does not apply.Was the patient¿s post-op care altered in any way due to the reported complication with the trocar? no.What is the current status of the patient? sure.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
(b)(4), date sent: 9/22/2022.D4: batch # unk.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample revealed that the 2d12lt device was returned with the housing cap component disassembled from the housing of the sleeve assembly.The housing cap and the housing are damaged.It should be noted that product failure is multifactorial.Based on the information currently available, a product issue was identified during the investigation of the sample received.This product issue will be addressed through ethicon¿s quality system.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.The manufacturing records couldn't be reviewed as the batch and lot number are unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPT DILATING 12X100 STABILITY
Common Device Name
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
329348013
MDR Report Key15245479
Report Number3005075853-2022-05378
Device Sequence Number6355525
Product Code GCJ
UDI-Device Identifier10705036014232
UDI-Public10705036014232
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K122512
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2D12LT
Device Catalogue Number2D12LT
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/25/2022
Supplement Date Received by Manufacturer08/30/2022
Initial Report FDA Received Date08/17/2022
Supplement Report FDA Received Date09/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-