ST. JUDE MEDICAL, INC.(CRM-SYLMAR) PROMOTE QUADRA CRT-D; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3239-40Q
Medical Device Problem Codes Over-Sensing (1438); Device Sensing Problem (2917)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/26/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr report product is for an ous event identified as same/similar during a retrospective review as a result of abbott¿s investigation.
 
Event or Problem Description
It was reported that the patient presented remotely via merlin.Net.Upon review of transmission, it was found that the implantable cardioverter defibrillator (icd) exhibited r-wave amplitude variation and myopotential over-sensing.Programming changes were made.There were no patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMOTE QUADRA CRT-D
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15245521
Report Number2017865-2022-19650
Device Sequence Number12347571
Product Code NIK
UDI-Device Identifier05414734507011
UDI-Public05414734507011
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
P030054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2013
Device Model NumberCD3239-40Q
Device Catalogue NumberEX1151
Device Lot Number3492293
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2021
Initial Report FDA Received Date08/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/12/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-