• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER; GBO CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER; GBO CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number N/A
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient required replacement of an ultrathane mac-loc locking loop multipurpose drainage catheter.It was noted at an unknown time post placement, the hub was separated from the catheter, not allowing chest tube to suction properly.The patient returned to interventional radiology for device exchange.No other adverse effects were reported for this incident.Additional information regarding the event and patient outcome has been requested but is currently unavailable.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.Reporter occupation: manager, supply chain services.Pma/510(k) #: exempt.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Additional Manufacturer Narrative
Additional information: this report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding patient and/or event details has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation: it was reported by a representative of (b)(6) hospital (florida, usa) that an ultrathane mac-loc locking loop multipurpose drainage catheter (rpn: ult14.0-38-25-p-6s-clm-rh; lot# 10073914) separated.The device was placed in the patient for use as a chest tube.Later, it was discovered that the mac-loc hub separated from the catheter which exposed the inner suture and prevented proper suction.A photo was provided showing the device separation.The physician was notified to assess the catheter.The patient returned to interventional radiology where the catheter was removed and replaced.No other adverse effects were reported due to this occurrence.Reviews of the complaint history, device history record (dhr), drawings, instructions for use (ifu), manufacturing instructions, quality control procedures, and specifications of the device, as well as a visual inspection of the customer provided photo, were conducted during the investigation.The complaint device was not returned.However, based on the provided photo from the customer, it was confirmed the proximal mac-loc adaptor separated from the catheter.Additionally, a document-based investigation evaluation was performed.In response to this incident, cook completed a review of the product device master record (dmr) and concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.The customer did not provide a lot number for this device.Therefore, cook medical inc.Performed an expanded sales search for the reported device and discovered a single lot, 10073914, sold to this facility.The dhr for lot 10073914 and related subassembly lot records no relevant non-conformances.The torque driver used in the manufacturing of this lot passed calibration.A further search of the reported lot number confirmed no additional complaints being received.Appropriate measures have been taken to address this failure mode.A capa was previously opened to further investigate this failure mode.Corrective actions have since been implemented following the manufacture of this device.Lot 10073914 was manufactured after corrective action implementation.Based on the dmr and dhr, there is no indication the device was manufactured out of specification.There is no evidence of nonconforming material in house or in the field.Cook reviewed the product labeling for the complaint device.The instructions for use(ifu) [ t_multi_rev5], multipurpose drainage catheter] states the following.Precautions: patients with indwelling drainage catheters should be evaluated routinely to ensure continuous function of the catheter.How supplied: "upon removal from package, inspect the product to ensure no damage has occurred." based on the information provided, inspection of the customer provided photo, and the results of the investigation, it was determined the cause of this event is related to component failure, unrelated to a design or manufacturing deficiency.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER
Common Device Name
GBO CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key15245528
Report Number1820334-2022-01354
Device Sequence Number6540934
Product Code GBO
UDI-Device Identifier00827002095045
UDI-Public(01)00827002095045(17)221010(10)10073914
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/10/2022
Device Model NumberN/A
Device Catalogue NumberULT14.0-38-25-P-6S-CLM-RH
Device Lot Number10073914
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/05/2022
Supplement Date Received by Manufacturer01/12/2023
02/21/2023
Initial Report FDA Received Date08/17/2022
Supplement Report FDA Received Date01/24/2023
03/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/10/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-