• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ANTHOLOGY SO TR NECK SZ 7-12; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ANTHOLOGY SO TR NECK SZ 7-12; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Model Number 71365707
Medical Device Problem Codes Degraded (1153); Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, both bumpers on the anthology so tr neck sz 7-12 and anthology ho tr neck sz 7-12 were either wearing off or completely worn off causing debris in set.As this was noticed in a non-surgical environment, there was not any patient involved.
 
Additional Manufacturer Narrative
The associated device was returned and evaluated.A visual inspection of the returned device confirms the magnet is worn and damaged causing the stated failure.This device exhibit signs of extreme wear and use.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.As the device is peeling away and it can no longer fit its purpose, the contribution of the device to the reported event could be corroborated.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTHOLOGY SO TR NECK SZ 7-12
Common Device Name
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15245623
Report Number1020279-2022-03717
Device Sequence Number5847362
Product Code JDG
UDI-Device Identifier03596010559845
UDI-Public03596010559845
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K123598
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71365707
Device Catalogue Number71365707
Device Lot Number05KM03196
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/11/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/01/2022
Supplement Date Received by Manufacturer09/14/2022
Initial Report FDA Received Date08/17/2022
Supplement Report FDA Received Date09/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-