ABBOTT VASCULAR RX HERCULINK ELITE STENT SYSTEM; STENT, RENAL Back to Search Results
Model Number 1011493-18
Medical Device Problem Codes Failure to Advance (2524); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/26/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the procedure was to treat a lesion located in the 70% stenosed, severely tortuous left renal artery.While advancing the 5.0x18mm herculink elite stent delivery system, the stent dislodged from the balloon.The stent delivery system was removed without issue and the stent successfully retrieved with a snare device.There was no adverse patient sequela or a clinically significant delay in procedure.Another same size device was used to complete the procedure successfully.No additional information was provided.
 
Additional Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Additional Manufacturer Narrative
A visual inspection was performed on the returned device.The reported stent dislodgement was confirmed.The reported failure to advance could not be evaluated as the exact anatomical conditions encountered by the device used during the procedure could not be replicated in the test laboratory.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents.The investigation determined the reported difficulties and subsequent treatments, including removal of foreign body, appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX HERCULINK ELITE STENT SYSTEM
Common Device Name
STENT, RENAL
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15246050
Report Number2024168-2022-08869
Device Sequence Number12347574
Product Code NIN
UDI-Device Identifier08717648078026
UDI-Public08717648078026
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P110001
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2023
Device Model Number1011493-18
Device Catalogue Number1011493-18
Device Lot Number0031661
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/26/2022
Supplement Date Received by Manufacturer10/03/2022
Initial Report FDA Received Date08/17/2022
Supplement Report FDA Received Date10/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/16/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-