Model Number 45-652050
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/20/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a splenic aneurysm embolization case, the coil couldn't go through the glidecath.When the physician pulled the coil out of the catheter, the coil detached inside.A new catheter was used and the procedure was successful.No harm or injury was reported.
 
Additional Manufacturer Narrative
A search for non-conformances associated with the reported part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was stated to be available for evaluation but has not yet been returned.If the device is received the investigation will be performed and a supplemental report will be submitted.The instructions for use (ifu) identifies premature coil detachment as a potential complication associated with use of the device.
 
Additional Manufacturer Narrative
The investigation of the returned coil system found the proximal connector kinked and the pusher damaged at the distal end.The implant was returned stretched, separated from the pusher, and stuck inside the microcatheter.The investigation found the pusher's monofilament with a tensile break shape at the tip which is consistent with the device experiencing excessive force that exceeded the strength of the monofilament causing the implant to separate from the pusher.The physical evaluation of the device could not identify the conditions or circumstances that led to the damage, but the damage is consistent with the device experiencing forces over specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZUR-35 DETACHABLE
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise
aliso viejo, CA 92656
MDR Report Key15246160
Report Number2032493-2022-00307
Device Sequence Number5847379
Product Code KRD
UDI-Device Identifier00810170016158
UDI-Public(01)00810170016158(11)210114(17)251231(10)2101145J8
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K151358
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2025
Device Model Number45-652050
Device Catalogue Number45-652050
Device Lot Number2101145J8
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/20/2022
Supplement Date Received by Manufacturer09/16/2022
Initial Report FDA Received Date08/17/2022
Supplement Report FDA Received Date10/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-