EXACTECH, INC. LOGIC TIBIA PS MOD INSRT SZ 2 15MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
02-012-35-2015 |
| Medical Device Problem Codes |
Unstable (1667); Naturally Worn (2988); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Discomfort (2330); Osteolysis (2377); Joint Laxity (4526); Implant Pain (4561); Swelling/ Edema (4577); Aseptic loosening (4883)
|
| Date of Event |
07/11/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Pending evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that his female patient underwent a left tkr at crystal clinic, inc.In akron, ohio, which is located in summit county.This was an optetrak logic device.Patient began to experience swelling in and around the left knee.Revision performed at crystal clinic on (b)(6) 2022.The polyethylene inserts used in the optetrak device were defective, leading to early aseptic loosening.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: the revision reported in the case may be due to prosthesis wear leading to osteolysis and loosening of implants.However, the prosthesis wear and contributions from implant positioning, instability, and/or patient-related factors to the sequence of events cannot be confirmed as the devices are not available for evaluation and limited clinical information was available at the time of evaluation.A contributing factor to the reported wear may have been have been result of the tibial insert being packaged in a non-conforming vacuum bag for more than five years.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b1, b2, d1, d4 (catalogue number), e1, e2, e3, g2, h4, h6 clinical codes, medical device problem codes, component code and types of investigation codes.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|