ABBOTT CONFIRM; IMPLANTABLE CARDIAC MONITOR Back to Search Results
Model Number DM2102
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/20/2022
Type of Reportable Event Malfunction
Event or Problem Description
During elective explant procedure, detachment of the header was observed.The device was explanted to resolve the event.The patient was in stable condition.
 
Additional Manufacturer Narrative
As received, device was visually inspected, and header of the device was found to be cracked and separated from the can due to excessive force by the user.This is consistent with improper handling of the device header by user.Device was received in backup mode with battery at end of service (eos) level due to normal usage.The implant duration exceeds the projected longevity indicating normal battery depletion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONFIRM
Common Device Name
IMPLANTABLE CARDIAC MONITOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15246604
Report Number2017865-2022-19317
Device Sequence Number15038206
Product Code MXC
UDI-Device Identifier05414734502573
UDI-Public05414734502573
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K133481
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 10/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2019
Device Model NumberDM2102
Device Catalogue NumberDM2102
Device Lot NumberS000064787
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/02/2022
Supplement Date Received by Manufacturer10/28/2022
Initial Report FDA Received Date08/17/2022
Supplement Report FDA Received Date10/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/11/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-