| Model Number |
M35054 |
| Medical Device Problem Codes |
Fracture (1260); Activation Failure (3270)
|
| Health Effect - Clinical Code |
Inadequate Pain Relief (2388)
|
| Date of Event |
07/28/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the control dial of the vaporizer "felt loose" and that the dosage of anesthetic agent was less than expected.Reportedly, the patient showed signs of beginning awareness.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation has just started; results will be provided in a follow-up report.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported aspect of compromised dial movement could initially been confirmed at the local dräger subsidiary.It was then decided to forward the vaporizer to the manufacturer for a potentially required in-depth factury repair.However, upon receipt the device was inspected and tested in the manufacturer's workshop but neither the compromized dial function nor the reported obervation of dosage deviation could be reproduced.Hence, a reliable explanation for the user's experience cannot be found.The delivery unit of the device which includes the mechanical components of the dial function has been replaced as a precautionary measure; the device was then returned to the customer in ready-to-use condition.
|
| |
|
Event or Problem Description
|
|
It was reported that the control dial of the vaporizer "felt loose" and that the dosage of anesthetic agent was less than expected.Reportedly, the patient showed signs of beginning awareness.
|
| |
|