• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONSOLIDATED MEDICAL EQUIPMENT COMPANY 1/4 IN ID X 12 FT L (.64CM IDX3.7M L) (20/CS); TUBE, ASPIRATING, FLEXIBLE, CONNECTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONSOLIDATED MEDICAL EQUIPMENT COMPANY 1/4 IN ID X 12 FT L (.64CM IDX3.7M L) (20/CS); TUBE, ASPIRATING, FLEXIBLE, CONNECTING Back to Search Results
Model Number 0036550
Medical Device Problem Code Delivered as Unsterile Product (1421)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
During incoming inspection, the distributor rejected this device, 0036550, for an insufficient heatseal.There was no contact with the patient as this was found during incoming inspection.This will be reported as a malfunction with potential for injury upon reoccurrence.
 
Additional Manufacturer Narrative
Received two 0036550 in original packaging.Lot number was verified.Performed a visual inspection of the device, there were no obvious signs of a breach in the seal however, the device was encroaching into the seal.Performed a functional inspection, the devices were dye leak tested which indicated that the packaging had an insufficient heat seal on one out of the two packages.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of 67 complaints, regarding (b)(4) devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be (b)(4).This issue will continue to be monitored through the complaint system to assure patient safety.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1/4 IN ID X 12 FT L (.64CM IDX3.7M L) (20/CS)
Common Device Name
TUBE, ASPIRATING, FLEXIBLE, CONNECTING
Manufacturer (Section D)
CONSOLIDATED MEDICAL EQUIPMENT COMPANY
ave. alejandro dumas 11321
complejo industrial chihuahua
chihuahua 31136
MX  31136
Manufacturer (Section G)
CONSOLIDATED MEDICAL EQUIPMENT COMPANY
ave. alejandro dumas 11321
complejo industrial chihuahua
chihuahua 31136
MX   31136
Manufacturer Contact
samantha dewberry
525 french road
utica, NY 13502
3152230184
MDR Report Key15248045
Report Number3007305485-2022-00128
Device Sequence Number6447773
Product Code BYY
UDI-Device Identifier10653405000525
UDI-Public(01)10653405000525(17)260426(10)202104274
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 08/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date04/26/2026
Device Model Number0036550
Device Catalogue Number0036550
Device Lot Number202104214
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/03/2022
Initial Report FDA Received Date08/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-