MAUDE Adverse Event Report: NATIONAL BIOLOGICAL CORPORATION PANOSOL 3D; PHOTOTHERAPY UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NATIONAL BIOLOGICAL CORPORATION PANOSOL 3D; PHOTOTHERAPY UNIT Back to Search Results
Model Number UVB-631-9
Health Effect - Clinical Code Partial thickness (Second Degree) Burn (2694)
Date of Event 07/22/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Patient reported to have 2nd degree burns after treatment with uv phototherpy equipment.Patient had previous treatment history at dermatologist's office using a clinical phototherapy unit.A home unit was prescribed to her as she no longer wanted to go to the office for treament due to covid concerns.She sustained the burns upon her first usage of the equipment.Altthough she said she had read the operations manual, she used a 4 minute exposure time on both the front and back sides of her body.The recommended first usage treatment time per the operations manual for a person with her skin type (type ii) was 1 min 13 sec.She had decided to use the same time she had been using at the dermatologist's office (5 minutes) minus 1 minute even though there is no correlation regarding treatment times between clinical and home units.The patient misinterpreted the manual that stated that the previous exposure time should be used if treatment was interrupted for 8-11 days.This was not intended to mean that the previous time using a non-like unit should be used.The patient provided photos of her burns to her dermatologist and was prescribed antibiotics, steroids and a topical cream for treatment.
 
Additional Manufacturer Narrative
Patient wanted to keep device, therefore it was never sent back to manufacturer for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PANOSOL 3D
Common Device Name
PHOTOTHERAPY UNIT
Manufacturer (Section D)
NATIONAL BIOLOGICAL CORPORATION
23700 mercantile rd.
beachwood OH 44122
Manufacturer (Section G)
NATIONAL BIOLOGICAL CORPORATION
23700 mercantile rd.
beachwood OH 44122
Manufacturer Contact
23700 mercantile rd.
beachwood, OH 44122
2168310600
MDR Report Key15248235
Report Number1521608-2022-00003
Device Sequence Number6390265
Product Code FTC
UDI-Device Identifier00816878020770
UDI-Public(01)00816878020770(11)220614
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K872650
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberUVB-631-9
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/22/2022
Initial Report FDA Received Date08/18/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/16/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age58 YR
Patient SexFemale
Patient EthnicityNon Hispanic
-
-