The customer reported, during lithotripsy, the main steel cord (not the basket) was broken.Due to the situation, the equipment could not be removed and the patient underwent laparotomy.A request for additional information is in progress.
|
|
This supplemental report is being submitted to provide additional information from the customer, the results of the legal manufacturer¿s investigation and the results of the device history records (dhr) review.New information added to the following fields: b5, h6, h10.H4: january 2022 based on the 3 digit lot information provided.The dhr with the lot number of the subject device was confirmed.No abnormalities detected in the dhr for the following items which related to the reported phenomenon.A definitive root cause was not identified.Based on the available information, the legal manufacturer determined the probable cause of the failure is likely due to the following: 1.Due to various factors such as the shape, numbers, hardness of the calculus, a force beyond the strength resistance was applied to the device during the lithotripsy.2.Due to the description above ¿1¿ description, the operating wire was broken.3.The lithotripsy was kept on going by combing emergency lithotripter and the subject device.However, a force beyond the strength resistance was applied to the operating wire due to various factors such as the shape, numbers, hardness of the calculus.4.Due to the description above ¿3¿ description, the operating wire was broken.Therefore, the basket wire part fell off into the patient¿s body.The instruction manual contains a warning to the user regarding this event.Contents of description (drawing number:gk5897, revision number : 20) do not use this instrument for a calculus that is assumed impossible to be crushed by a lithotriptor.The pipe or the basket wire may break and part of this instrument may remain in the body.Use this instrument by having the settings to switch to open surgery and the hospitalization plan ready in case the calculus cannot be crushed by lithotriptor bml-110a-1.A lithotriptor cannot always crush all calculi captured in the basket.Operation of this instrument is based on the assumption that open surgery is possible as an emergency measure.If the calculus is too hard, it is possible that the damages shown in chapter 5, ¿emergency treatment¿ may occur.Use the lithotriptor by considering that it may lead to damaging the instrument and that open surgery may have to take place.This instrument will deform and/or deteriorate by performing lithotripsy.When lithotripsy is repeated, it will deform and/or deteriorate furthermore.By such deformation and/or deterioration, calculus may not be crushed and/or the instrument with calculus engaged may not be removed from the body.If lithotripsy is required to be repeated in a single case, make sure to check each time that no abnormality is found in action and/or appearance (e.G.Basket wire cut or worn, tube sheath bent, notable coil sheath bent or gap etc.).Stop use when any abnormality is detected.During lithotripsy, keep the portion from the coil sheath to the bml handle straight in line with the scope¿s biopsy valve, as much as possible.If not straight, the coil sheath may bend, calculus may not be crushed, and/or the instrument with calculus engaged may not be removed from the body.Do not rotate the bml handle knob abruptly.This instrument may break, and/or calculus may not be crushed.Also, the instrument with calculus engaged may not be removed from the body.
|