• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. INTERCHANGEABLE HUMERAL ASSEMBLY; PROSTHESIS, ELBOW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. INTERCHANGEABLE HUMERAL ASSEMBLY; PROSTHESIS, ELBOW Back to Search Results
Model Number 32-8105-25-04
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Failure of Implant (1924); Unspecified Infection (1930)
Date of Event 07/29/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Concomitant medical products: 32810505301 62875268 interchangeable ulnar assembly for cemented use only small left.Report source: foreign- australia.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device was requested but not returned by the hospital.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the patient had an initial elbow surgery.Subsequently, the patient underwent revision procedure approximately seven years post initial surgery due to infection and loosening of humeral component.All components were removed and replaced.No additional patient consequences were reported.
 
Event or Problem Description
No additional event information is available at the time of this report.
 
Additional Manufacturer Narrative
Proposed g code: mechanical (g04) - stem.Reported event was unable to be confirmed due to limited information received from the customer.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERCHANGEABLE HUMERAL ASSEMBLY
Common Device Name
PROSTHESIS, ELBOW
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15249232
Report Number0001822565-2022-02320
Device Sequence Number5905648
Product Code JDC
UDI-Device Identifier00889024508590
UDI-Public(01)00889024508590(17)190831(10)62748293
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K181307
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2019
Device Model Number32-8105-25-04
Device Catalogue Number32810502504
Device Lot Number62748293
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/29/2022
Supplement Date Received by Manufacturer12/02/2022
Initial Report FDA Received Date08/18/2022
Supplement Report FDA Received Date12/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/28/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age61 YR
Patient SexMale
-
-