DEPUY ORTHOPAEDICS INC US SIG RP AOX CRV INS SZ2.5 10; SIGMA KNEE PRIMARY : KNEE TIBIAL INSERT Back to Search Results
Model Number 1961-92-021
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Erythema (1840); Fever (1858); Unspecified Infection (1930); Pain (1994); Discomfort (2330); Fluid Discharge (2686); Swelling/ Edema (4577)
Date of Event 07/06/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Clinical adverse event received for deep infection and pain.Event is serious and is considered severe.Event is possibly related to both device and procedure.Date of implant: (b)(6) 2022; date of event: (b)(6) 2022; (right knee).Treatment: revision; insert was removed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary: no device associated with this report was received for examination.Available xray evidence was reviewed.Based in attached photographic evidence, it cannot be determinate a depuy's device failure or malfunction which could contribute to reported event depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation may be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Event or Problem Description
Medical records received: primary operative notes (b)(6) 2022 indicate the patient received a right total knee replacement due to end stage osteoarthritis.The surgery was completed without indication of complication by the surgeon.Revision operative notes (b)(6) 2022 indicate the patient received a right total knee revision due to pain, discomfort, swelling, fever, wound draining, erythema, warmth and infection of the right knee.Upon entering the knee, gross pus was encountered, and a thorough irrigation and debridement scrub was performed as well as inert exchange.The surgery was completed without indication of complication by the surgeon.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Additional information received indicated that a clinical adverse event was received for continued infection/deep infection.Laboratory tests were provided to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIG RP AOX CRV INS SZ2.5 10
Common Device Name
SIGMA KNEE PRIMARY : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key15249320
Report Number1818910-2022-16079
Device Sequence Number6355349
Product Code NJL
UDI-Device Identifier10603295076124
UDI-Public10603295076124
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1961-92-021
Device Catalogue Number196192021
Device Lot Number3647944
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/03/2022
Supplement Date Received by Manufacturer09/22/2022
03/01/2023
Initial Report FDA Received Date08/18/2022
Supplement Report FDA Received Date09/23/2022
03/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DEPUY CMW 2 20G; DEPUY CMW 2 20G; MBT CEM KEEL TIB TRAY SZ2.5; MBT CEM KEEL TIB TRAY SZ2.5; PFC*SIGMA/OV/DOME PAT 3PEG,35; PFC*SIGMA/OV/DOME PAT 3PEG,35; SIG C/R POROCOAT FEM RT SZ 2.5; SIG C/R POROCOAT FEM RT SZ 2.5; SIG RP AOX CRV INS SZ2.5 10; SIG RP AOX CRV INS SZ2.5 12.5
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
Patient SexFemale
Patient Weight95 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-