Model Number 1365-05-000
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Numbness (2415); Metal Related Pathology (4530)
Date of Event 08/03/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that patient had a pinnacle mom hip replacement.Patient began experiencing numbness in his feet.Surgeon tested for metal ion levels and stated one was elevated while the other was in normal ranges.The decision was made to do a head liner exchange from mom to ceramic on poly.A head was removed along with pinnacle mom liner.Surgeon replaced with a ceramic and a liner uneventfully.No patient harm was noted during the procedure.There was no surgical delay.Doi- unknown; dor- (b)(6) 2022; affected side- left hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.Available xray evidence was reviewed.Based in attached photographic evidence, it cannot be determinate a depuy's device failure or malfunction which could contribute to reported event.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation may be re-opened as necessary.Device history lot : a manufacturing records evaluation (mre) was not performed as no lot number was provided for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
12/14 ARTICUL 40MM M SPEC+1.5
Common Device Name
ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY INT'L LTD. 8010379
st anthonys road
leeds LS11 8 DT
UK   LS11 8 DT
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key15249658
Report Number1818910-2022-16089
Device Sequence Number6447525
Product Code LPH
UDI-Device Identifier10603295032960
UDI-Public10603295032960
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K060031
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1365-05-000
Device Catalogue Number136505000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/03/2022
Supplement Date Received by Manufacturer09/29/2022
Initial Report FDA Received Date08/18/2022
Supplement Report FDA Received Date09/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
12/14 ARTICUL 40MM M SPEC+1.5.; UNKNOWN FEMORAL STEM.; UNKNOWN HIP ACETABULAR LINERS.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient SexMale
-
-