| Model Number |
BF-XP60 |
| Medical Device Problem Codes |
Contamination (1120); Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is returned and an evaluation completed for it.Upon inspection of the device, it was observed that the distal end was dirty with foreign material around the biopsy channel opening.This is likely due to failure to clean adequately.Other observations for the device are adhesive of the distal end rubber coating (a-rubber) is attached.The blurry image issue could not be duplicated.Evaluation is ongoing.Supplemental report(s) will be submitted when any relevant new information is available.
|
| |
|
Event or Problem Description
|
|
Customer returned the device for evaluation and repair of a blurry image issue noted during preparation for use.There is no patient involvement and no harm reported to any patient.Upon evaluation of the returned device, it was observed that the distal end was dirty with foreign material around the biopsy channel opening.This medwatch is being submitted for the issue of foreign material around the biopsy channel opening as observed during device evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information has been received for this event.This supplemental report is being submitted to provide this information.There was no adequate reprocessing/brushing performed for the device at the time of return.General information for reprocessing performed at the facility: pre-cleaning is performed immediately after a procedure.Steps followed for pre-cleaning are wiping the outer surfaces of scope and aspiration of suction channel with water and detergent solution and air.Detergent used for pre-cleaning is 3m.Brushing of the endoscope channels is sometimes missed during manual cleaning.When performed, the brush used is reusable olympus bw-15b.Cleaning and disinfectant solutions used with the endoscope are johnson and johnson 3m and ga (cidex).Customer does not use automatic endoscopy reprocessor (aer).Endoscopes are dried with wiping and air suction and stored in an endoscope almirah cabinet after reprocessing.
|
| |
|
Event or Problem Description
|
|
Addendum feb 7, 2022: intended procedure is diagnostic bronchoscopy.Procedure was completed using similar olympus device and with no substantial delay.There was no impact on the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the foreign material coming out of forceps channel was observed to be a brown solid, but otherwise, could not be identified.The root cause of the issue could not be determined, however, based on information provided by the customer, the device was not properly reprocessed/brushed when sent to olympus and brushing of the channels during manual cleaning is sometimes forgotten.Therefore, it is probable that foreign matter remained in the channel due to the customers reprocessing deviation from the device instructions for use.The event can be prevented by following the instructions for use which state: ¿operation manual: 5.3 returning the endoscope for repair: thoroughly clean and high-level disinfect or sterilize the endoscope before returning it for repair.Improperly reprocessed equipment presents an infection control risk to each person who handles the endoscope within the hospital or at olympus.Reprocessing manual: 1.4 precautions: all channels of the endoscope must be cleaned and high-level disinfected or sterilized during every reprocessing cycle, even if the channels were not used during the previous patient procedure.Otherwise, insufficient cleaning and disinfection or sterilization of the endoscope may pose an infection control risk to the patient and/or operators performing the next procedure with the endoscope¿.Olympus will continue to monitor field performance for this device.
|
| |
|