MEDTRONIC PUERTO RICO OPERATIONS CO. SURESCAN; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
|
Back to Search Results |
|
Model Number 97714 |
Device Problem
Battery Problem (2885)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
Information was received from a patient who was implanted with an implantable neurostimulator (ins) for unknown indications for use.The reason for call was pt reported they were seeing the por message display on their programmer for almost a year.Pt inquired on how to get their device into mri mode and the warning por message displayed.Caller (nurse in the background) noted that the pt was seeing the por message when attempting to charge (ps understood this as their implant).Pt stated they were seeing the por message for a long time and pt was in contact with them for over a year.Pt stated they had an appointment set up for april and had a pain clinic appointment in july.Pt noted they were trying for many months to get their device restarted.The patient was redirected to their healthcare provider to further address the issue.Ps had difficulty hearing the pt during the call as the pt's connection kept cutting off.Ps advised pt to connect with a representative and pt noted the representatives did not show up to their appointment.Pt also noted they had an appointment at the bone and joint clinic this friday and next friday.Redirected caller: patient's hcp ps emailed representatives for visibility.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from the patient.They reported that the cause of the por was that the battery had stayed discharged for too long.The manufacturer representative (rep) was able to remove the por, and resolved the issue.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|