• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES NAVIO SURGICAL SYSTEM US; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES NAVIO SURGICAL SYSTEM US; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number NPFS02000
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fall (1848); Bone Fracture(s) (1870)
Date of Event 02/05/2020
Type of Reportable Event Serious Injury
Event or Problem Description
On the literature review "initial safety profile assessment of the navio robotic-assisted total knee arthroplasty", it was reported that, after undergoing a navio-assisted ra-tka procedure, one (1) patient experienced a fall resulting in a peri-prosthetic fracture.It is unknown how this adverse event was treated.No further information is available.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).Article cite: initial safety profile assessment of the navio robotic-assisted total knee arthroplasty, international society for technology in arthroplasty (ista) meeting, 32nd annual congress, toronto, canada, october 2019.
 
Additional Manufacturer Narrative
The navio surgical system us, part number npfs02000, intended for treatment was not returned for evaluation.A relationship between the reported event and the device could not be established.The clinical/medical investigation concluded that without the requested medical documentation, definitive clinical factors which could have contributed to the reported event could not be concluded.The patient impact beyond the reported fall and subsequent periprosthetic fracture along with the clinical outcome could not be determined based on the limited information provided in the literature.A complaint history review for similar reported/confirmed complaints has identified prior events.While all products meet required manufacturing specifications prior to release a serial number, lot number, part revision or software version is required to link the device to a dhr or nc investigation.As with any surgical procedure, there is risk involved.Potential complications accompanying surgery may occur, including: allergic reaction (anaphylactic and minor), infection, mild to serious physical injury, localized static shock, delay in the operation, surgical site nerve injury, vascular injuries of the lower extremity, soft tissue damage, major bone gouging at the surgical site, bone fracture, immature implant failure, unstable knee joint, limited or restricted knee range of motion, major blunt impact injury, unintended laceration/puncture wound, and osteonecrosis.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still within the product risk profile.No reasonable contributing factors could be identified based on the received complaint information and investigation results.Based on the investigation, the need for a corrective action is not recommended or required at this time.Should any additional information be received the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO SURGICAL SYSTEM US
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15251338
Report Number3010266064-2022-00569
Device Sequence Number5681577
Product Code OLO
UDI-Device Identifier00885556628416
UDI-Public885556628416
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Study,Literature,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberNPFS02000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/01/2022
Supplement Date Received by Manufacturer11/17/2022
Initial Report FDA Received Date08/18/2022
Supplement Report FDA Received Date11/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-