MAUDE Adverse Event Report: ZOLL CIRCULATION INC ZOLL IVTM ICY CATHETER; CENTRAL VENOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION INC ZOLL IVTM ICY CATHETER; CENTRAL VENOUS CATHETER Back to Search Results
Model Number IC-3893
Medical Device Problem Codes Fluid/Blood Leak (1250); Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/27/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported complaint of a leak on the icy catheter (lot #173639) was confirmed during functional testing.A bonding leak was observed at distal end of proximal balloon during the functional pressure leak test.The probable root cause could be a latent defect at the bond.Visual inspection of the returned catheter was performed, and no physical damage was found on the catheter.Blood residue was observed in the balloons and luered tubings.Functional leak test of the returned catheter was performed, and all infusion ports and lumens were flushed without resistance.During the functional pressure leak test, the catheter was connected to a pressurized inflation device.Upon pressurizing the catheter up to 100 psi, a bonding leak was observed at distal end of proximal balloon; thus, confirming the reported complaint.It is unlikely that the defective catheter was shipped.During manufacturing, all catheters are 100% inspected for leaks by subjecting them to pressure testing.Only units that passed are moved to the next process.Historical complaints were reviewed for information related to the reported complaint, and there were no similar complaints reported for the icy catheter with lot number 173639.
 
Event or Problem Description
On (b)(6) 2022, ivtm therapy began.On (b)(6) 2022, the saline bag was observed to be almost empty and the saline solution in the air trap was also decreased.After checking that no saline was leaking from around the console, the customer determined that the icy catheter (lot #173639) was the cause of the saline leak and replaced the catheter.The start-up kit (suk) (lot # 171714) was replaced just in case.No device malfunction was reported for the suk.After replacing the icy catheter and suk, ivtm therapy continued with the same console without any problems.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZOLL IVTM ICY CATHETER
Common Device Name
CENTRAL VENOUS CATHETER
Manufacturer (Section D)
ZOLL CIRCULATION INC
2000 ringwood ave
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION INC
2000 ringwood ave
san jose CA 95131
Manufacturer Contact
kim thoa nguyen
2000 ringwood ave
san jose, CA 95131
4084192922
MDR Report Key15252215
Report Number3010617000-2022-01272
Device Sequence Number14904883
Product Code NCX
UDI-Device Identifier00849111075527
UDI-Public00849111075527
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K101987
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIC-3893
Device Catalogue Number8700-0782-03
Device Lot Number173639
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/08/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/27/2022
Initial Report FDA Received Date08/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/11/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-