• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE VIDEO PROCESSOR/ILLUMINATOR (VPI); LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE VIDEO PROCESSOR/ILLUMINATOR (VPI); LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number PC9000
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/04/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device lost image and powered off during a procedure.The users were unable to get the device to power back on, which resulted in a 15 minute delay while a replacement was brought in.The procedure was completed successfully and there was no report of adverse consequences.
 
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that the device lost image and powered off during a procedure.The users were unable to get the device to power back on, which resulted in a 15 minute delay while a replacement was brought in.The procedure was completed successfully and there was no report of adverse consequences.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: can not bootup probable root cause: electrostatic discharge.Cooling system (vpi) failure.Software error/failure.Camera overheating.Camera malfunction.Dropped camera.The reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIDEO PROCESSOR/ILLUMINATOR (VPI)
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
burnaby
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
burnaby
san jose CA 95138
Manufacturer Contact
chester rebugio
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key15252529
Report Number3012345110-2022-00004
Device Sequence Number5745677
Product Code GCJ
UDI-Device Identifier00858701006049
UDI-Public00858701006049
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K150956
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician Assistant
Type of Report Initial,Followup
Report Date (Section B) 10/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPC9000
Device Catalogue NumberPC9001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/11/2022
Supplement Date Received by Manufacturer08/11/2022
Initial Report FDA Received Date08/18/2022
Supplement Report FDA Received Date10/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-