| Model Number |
TSVMWH11 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
06/22/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint number (b)(4).Pma/510k: additional 510(k) number is k133339.
|
| |
|
Event or Problem Description
|
|
No event was reported by the customer however the implant was received fractured at the implants head.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimmerbiomet complaint number (b)(4).The reported device was returned for investigation.Visual evaluation of the returned product identified that the implant fractured around collar/platform.No pre-existing conditions were noted on the product experience report (per).The reported device tooth location was not reported.The implant was used for approximately 3 years, 6 months.The customer did not provide any pictures or x-rays.Complainant reported: no event was reported by the customer however the implant was received fractured at the implants head.A fractured implant was confirmed.A definitive root cause for this complaint could not be determined.A device history record (dhr)review was performed and no related deviations or nonconformances were noted.Also, a complaint history search was performed using our complaint handling system and there were no related complaints for this product lot.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|