• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. A.T.S. 4000TS TOURNIQUET W/HOSES AND LOP SENSOR (PERSONALIZED PRESSURE®); SURGICAL PNEUMATIC TOURNIQUET SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. A.T.S. 4000TS TOURNIQUET W/HOSES AND LOP SENSOR (PERSONALIZED PRESSURE®); SURGICAL PNEUMATIC TOURNIQUET SYSTEM Back to Search Results
Model Number N/A
Medical Device Problem Codes Deflation Problem (1149); Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/21/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The product has been returned to zimmer biomet, and the investigation is in process.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
It was reported that device failed leak test.There was spontaneous deflation during a leak test that was performed during preventative maintenance.There was no patient involvement.It was reported that no further information is available.
 
Additional Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(6).This medwatch is being filed to relay additional information.The following sections were updated/corrected: b4, b5, d4, g3, g6, h2, h3, h4, h6, h10.Review of the most recent repair record determined the unit was confirmed to be failing the leak test.The slow deflate valve on the main manifold was and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional event information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
A.T.S. 4000TS TOURNIQUET W/HOSES AND LOP SENSOR (PERSONALIZED PRESSURE®)
Common Device Name
SURGICAL PNEUMATIC TOURNIQUET SYSTEM
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key15254363
Report Number0001526350-2022-00801
Device Sequence Number20285580
Product Code KCY
UDI-Device Identifier00889024376731
UDI-Public(01)00889024376731(11)160219(10)63278368
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K123553
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number60400010100
Device Lot Number63278368
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/12/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/08/2022
Supplement Date Received by Manufacturer09/02/2022
Initial Report FDA Received Date08/18/2022
Supplement Report FDA Received Date09/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/20/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-