• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD (SUZHOU) BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD (SUZHOU) BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 383954
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/27/2022
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that the bd pegasus¿ safety closed iv catheter system leaked medication from the tubing.The following information was provided by the initial reporter, translated from chinese: "the head nurse of the operating room reported that during the use of the product for the patient, the extension tube was found to come out of the y-shaped interface, resulting in drug liquid leakage, and the patient was dissatisfied.".
 
Manufacturer Narrative
The following fields were updated due to additional information: d.9.Device available for eval?: yes.D.9.Returned to manufacturer on: 18-aug-2022.Investigation summary: in response to the event reported, a device history review was conducted for lot number 1293369.Our records show that this is the only instance of this issue occurring in this production batch.According to the sampling plan applied for product performance, this lot was accepted and released without defects being noted during the final assembly or visual inspections.Additionally a sample was returned to our facility which displayed an extension tube that had been separated from the y-shaped housing.The issue has been confirmed.Based on the visual analysis of the device, our engineers have determined that the root cause for this event is related to the swaging process.Swaging is implemented to ensure that the cavity which houses the tubing and metal cuff is of an optimized depth.A shallow cavity can result in the tubing becoming displaced from the housing.To prevent future occurrences of this event our engineers have issued a retraining to the relevant personnel.Complaints received for this product and condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business regularly reviews the collected data for identification of emerging trends.
 
Event Description
It was reported that the bd pegasus¿ safety closed iv catheter system leaked medication from the tubing.The following information was provided by the initial reporter, translated from chinese: "the head nurse of the operating room reported that during the use of the product for the patient, the extension tube was found to come out of the y-shaped interface, resulting in drug liquid leakage, and the patient was dissatisfied".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BD (SUZHOU)
no. 5 baiyu road
suzhou industrial park
suzhou
Manufacturer (Section G)
BD (SUZHOU)
no. 5 baiyu road
suzhou industrial park
suzhou
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15254677
MDR Text Key305266599
Report Number3014704491-2022-00367
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/18/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number383954
Device Lot Number1293369
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/09/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/20/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-