• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. TRULIANT PIN PULLER SYRINGE STYLE; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. TRULIANT PIN PULLER SYRINGE STYLE; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number TRULIANT PIN PULLER SYRINGE STYLE
Medical Device Problem Code Component Misassembled (4004)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Event or Problem Description
It was reported that during a procedure, a piece from the pin puller fell into the wound site and was immediately removed by the surgeon.There was no surgical delay.The patient was last known to be in stable condition following the event.The product is available for return.
 
Additional Manufacturer Narrative
H6: investigation determined the broken truliant pin puller reported, was likely the result of: 1) the breakage area of the component being relatively thin which does not offer great safety margin when the instrument is subject to off-axis loading.2) off-axis loading when the user applies a bending moment to the device during pin driving/pulling.Corrective action taken: 1) the design was updated to improve the strength of the breaking area.2) the operative technique was updated to instruct users to avoid using the instrument offaxis.3) a new short-headed pin and an updated syringe pin puller was designed to address the sticking with the baseplate trial while not compromising the strength of the syringe pin puller at the previously broken area.
 
Additional Manufacturer Narrative
Upon reassessment of the reported event, it was determined that the event did not involve a death or serious injury, nor would be likely to cause or contribute to a death or serious injury if the reported malfunction were to recur.As a result, this complaint event is no longer considered reportable by exactech and this report may be disregarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRULIANT PIN PULLER SYRINGE STYLE
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66 court
gainesville, FL 32653
3523771140
MDR Report Key15254746
Report Number1038671-2022-00943
Device Sequence Number14889561
Product Code LXH
UDI-Device Identifier10885862585882
UDI-Public10885862585882
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
UNK
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTRULIANT PIN PULLER SYRINGE STYLE
Device Catalogue Number02-029-90-4100
Device Lot Number115977007
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/20/2022
Supplement Date Received by Manufacturer01/10/2023
12/11/2025
Initial Report FDA Received Date08/18/2022
Supplement Report FDA Received Date02/07/2023
12/11/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-